GPU-based Monte Carlo source modeling and simulation for radiation therapy involving Varian TrueBeam LINAC

Loading...
Thumbnail Image
Authors
Lin, Hui
Issue Date
2018-08
Type
Electronic thesis
Thesis
Language
ENG
Keywords
Nuclear engineering and science
Research Projects
Organizational Units
Journal Issue
Alternative Title
Abstract
To achieve this goal, a source model of the Varian TrueBeam LINAC and an efficient, First-Compton-based source particle sampling method are developed and are integrated with a fast coupled electron-photon transport module of the recently developed high-performance Monte Carlo code, ARCHER. The simulation results of different parts of the source model are verified with established dose calculation codes and validated with experimental measurements. Isodose distributions, dose difference maps and three-dimensional Gamma index tests are performed to demonstrate and quantify the benchmark results. To demonstrate the capability of the platform in clinical applications, three clinical treatment plans (the lung, breast and prostate) are simulated and analyzed. The hardware devices involved in this study include one NVIDIA TitanX GPU, one NVIDIA K40 GPU, one NVIDIA K20 GPU and one 6-cores Intel X5650 CPU. From the perspective of performance, the entire simulation of each clinical plan, starting from the source modeling to the end of particle transport within the patient phantom, can be finished in less than one minute using a single Titan X GPU card with 1% statistical error. In comparison, the same cases take the EGSnrc code hundreds of CPU hours to finish. The presented results indicate that this new version of ARCHER code was successfully developed and tested as a clinically deployable dose engine for external photon beam IMRT plans in the clinical setting.
Description
August 2018
School of Engineering
Full Citation
Publisher
Rensselaer Polytechnic Institute, Troy, NY
Terms of Use
Journal
Volume
Issue
PubMed ID
DOI
ISSN
EISSN